11.4. Status messages / error messages
The programmable controller sends data (including error messages) only when the computer
sends a query.
Status messages Description
00 MANUAL STOP Programmable controller in "OFF" state.
01 MANUAL START Programmable controller in keypad control
mode.
02 REMOTE STOP Programmable controller in "r OFF" state.
03 REMOTE START Programmable controller in remote control
mode.
Error messages Description
-02 REFRIGERATOR ALARM Control cable of the refrigerated circulator or
MVS solenoid valve controller short-circuited or
interrupted.
-03 EXCESS TEMPERATURE WARNING
High temperature warning "
".
-04 LOW TEMPERATURE WARNING
Low temperature warning "
".
-05 WORKING SENSOR ALARM Working temperature sensor short-circuited or
interrupted.
-07 I
2
C-BUS ERROR Internal error when reading or writing the I
2
C
bus.
-08 INVALID COMMAND Invalid command.
-09 COMMAND NOT ALLOWED IN CURRENT
OPERATING MODE
Invalid command in current operating mode.
-10 VALUE TOO SMALL Entered value too small.
-11 VALUE TOO LARGE Entered value too large.
-12 TEMPERATURE MEASUREMENT ALARM Error in A/D converter.
-13 WARNING : VALUE EXCEEDS
TEMPERATURE LIMITS
Value lies outside the adjusted range for the
high and low temperature warning limits. But
value is stored.
-14 TEMPERATURE/LEVEL ALARM
Safety temperature alarm
-15 EXTERNAL SENSOR ALARM External control selected, but external Pt100
sensor not connected.
-16 TRIAC/RELAY CONNECTION OPEN Heating circuit interrupted.
-17 TRIAC SHORTED Heating circuit short-circuited.
-18 RELAY SHORTED Defective alarm relay.