Command
Parameter
Response of programmable controller
in_mode_04
none Internal/external temperature control:
0 = Temperature control with Pt100 sensor „INT“.
1 = Temperature control with Pt100 sensor „EXT“.
in_mode_05
none Circulator in Stop/Start condition:
0 = Stop
1 = Start
in_par_01
none Time constant of the external bath.
in_par_02
none Internal slope.
in_par_03
none Time constant of the internal bath.
in_par_04
none Band limiting (max. difference between the temperatures in
the internal bath and external system).
in_par_05
none Ratio for max. cooling power versus max. heating power.
in_par_06
none Xp control parameter of the internal controller.
in_par_07
none Tn control parameter of the internal controller.
in_par_08
none Tv control parameter of the internal controller.
in_par_09
none Xp control parameter of the cascade controller.
in_par_10
none Proportional portion of the cascade controller.
in_par_11
none Tn control parameter of the cascade controller.
in_par_12
none Tv control parameter of the cascade controller.
out commands: Setting parameters or temperature values.
Command
Parameter Response of circulator
out_mode_01
0 Use working temperature "T1"
out_mode_01
1 Use working temperature "T2"
out_mode_02
0 No identification.
Temperature control by using the stored parameters.
out_mode_02
1 Single identification of controlled system after the next start.