Electrical connections
42
1
2
4
5
REG+E-PROG
Programmer input / temperature recorder output (31)
Analog inputs / outputs see page 35
Pin Signal
1 Voltage output Channel 1 0 ... 10 V
2 Voltage output Channel 2 0 ... 10 V
3 GND for outputs 0 V
4 Programmer input EPROG 0 to 10 V / 0 to 20 mA
5 Current output Channel 3 0 to 20 mA / 4 to 20 mA
6 GND for Progammer 0 V
13
2
4
5
Control output (32)
The
connector may be used for control of JULABO
refrigerated programmable controllers or as output for alarm
messages.
Pin assignment:
Pin Signal
1 +24 V (I max. current 25 mA)
2 0 V
3 Alarm relay
4 Reserved - do not use!
5 Cooling pulse
Circuit: Operation = relay powered
Alarm = relay not powered
ALARM
Alarm output (34)
(for external alarm signal)
This potential-free change-over contact is activated in case of
an alarm when pins 2 and 3 are connected.
Switching capacity max. 30 W / 40 VA
Switching voltage max. 125 V/
Switching current max. 1 A
1
3
2
SF
INT
EXT
Socket for temperature sensors (19, 20, 21)
Pin assignment:
Pin Signal
1 Current+
2 Voltage+
3 Voltage-
4 Current-