Julabo LC 4 Temperature controller Manual de usuario Pagina 25

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 35
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 24
LC4
25
For each parameter set, the control parameters Xp, Tn and Tv
may be individually modified and stored (see page 19).
Preference:
Quick access to parameters determined during previous
applications.
Factory-preset parameter sets for the following
controlled processes:
Set 1: Xp = 4 °C
Tn = 150 s
Tv = 10 s
Directly heated liquid bath
Filling volume: 4.5 litres of water
Setpoint temperature: 70 °C
Heater capacity: 2000 W
Set 2 Xp = 10 °C
Tn = 300 s
Tv = 25 s
Directly heated liquid bath
Filling volume: 4 litres of Thermal H
Setpoint temperature: 150 °C
Heater capacity: 2000 W
Set 3 Xp = 20 °C
Tn = 1800 s
Tv = 165 s
Heating hood model JH2000 with 2000 ml round bottom
flask
Filling volume: 2 litres of water
Setpoint temperature: 70 °C
Heater capacity: 480 W (stage III)
Set 4 Xp = 35 °C
Tn = 1800 s
Tv = 130 s
Heating hood JM500 with 500 ml round bottom flask
Filling volume: 0.5 litres of water
Setpoint temperature: 70 °C
Heater capacity: 150 W (stage II)
Set 5 Xp = 15 °C
Tn = 1550 s
Tv = 120 s
Heating hood JM500 with 500 ml round bottom flask
Filling volume: 0.5 litres of Thermal H
Setpoint temperature: 100 °C
Heater capacity: 150 W (stage II)